This system has an intuitive, user-friendly interface and allows you to program up to 50 assay protocols, plot and store standard curves and calculate concentrations. You can connect to a PC or directly to a printer to retrieve data.

Specifications | |
---|---|
Operating voltage | 24 VDC (plus an internal contact) |
Power consumption | < 30 VAC |
Set-up location | The instrument may be operated in rooms having low humidity |
Safety standards | EN 61010-1, EN 61326-1, EN 61000-3-2, EN 61000-3-3 |
Temperature range | Storage: 0º - 40ºC Operation: 15º - 35ºC |
Altitude | < 2000 above sea level |
Humidity | 10 – 85% no condensation |
Degree of pollution | Class 2 |
Dimensions (W x D x H) | 330 x 430 x 270 mm |
Weight | 17kg |
Detection unit | Low-noise photomultiplier with single photon counting technology (LB 9606) |
Measurement technology | Luminescence |
Sensitivity | < 0.001 mIU/L (TSH) |
Dynamic range | > 6 decades |
Cross talk | Low cross talk due to cross talk reduction methods: < 10-6 |
Plate formats | All 96 well microplates with external dimensions: 86.0 x 128.2 x 14.6 x mm (W x D x H); height variability +0.4 mm, -0.3mm |
Operation modes | Measuring time: 0.1-200 seconds Measuring direction: by row or by column Shaking: 3 modes, variable amplitude and speed Delay: up to 600 seconds |
Data evaluation | Evaluation Qualitative cut-off determination with up to 3 categories. Immunometric (ILMA) and competitive (LIA) measurements with standard curve calculation, master curves, reference curves, and curve fitting Export Output of all raw data and calculations in matrix or list format. |
Result display | Raw data in matrix form. All other data in spreadsheet form. With Full Standardization the curve is printed out graphically. Integrated statistics function. bOutput of all measurement data to printer. |
Ports | USB, RS232 (9 pin), parallel (25 pin) |